Detection Method: Absorbance
Measurement Modes: Endpoint and Kinetic
Scan function: High resolution scanning. Up to 29 points/well
Light Source: LED Calibration: Self calibrating
Plate Type: 96 (8x12), 48 (4x12) flat-, U- and V-bottom
Wavelength Range: 400nm - 700nm
Measurement channels: 8
Indication Range: 0.000 4.500 OD
Resolution: 0.001 OD (Indication increment)
Accuracy:<±0.5% at 1.0 OD (492 nm)
Linearity: <±0.75% and ± 0.005 OD from 0.1- 3.0 OD
<±1.5% and ± 0.005 OD from 3.0- 3.5 OD
Repeatability: <0.15% at 1.0 OD
Measurement time: 5 sec single wavelength, 8 sec dual wavelength (kinetic interval)
Filters: up to 8. Standard 405nm, 450, 492, 620
Shaking: 3 modes
Dimensions: 33.5cm x 20.5cm x 47.0 cm (w x h x l)
Weight: ~13.0kg net
Power supply: external
voltage range: 100-240VAC ±10%, 47-63Hz
Data connections:Serial Interface RS232 (9-pin), USB 2.0
Operating: temperature: 10°C to 40°C
humidity: 15%-85% RH non-condens.
Storage: temperature: -10°C to 50°C
humidity: <95% RH non-condensing
Scope of supply:
Instrument with 4 filters (405, 450, 492,
620nm), power supply, serial cable, USB cable, AUTOsoft
Control and Evaluation Software (Windows ) and user manual on CD
Other filters: on request